MBMap Registry
  1. Devices
  2. Seeed Studio
  3. SenseCAP UV Radiation Sensor S-ZW-01A

Seeed Studio SenseCAP UV Radiation Sensor S-ZW-01A Modbus register map

Also covers 101991048

Industrial UV radiation sensor, 240 to 395 nm, RS-485 (Modbus RTU)

  • Official
  • 2 registers
  • Modbus RTU
  • Revision 1 · 2026-10-18

Official: Written from the manufacturer's documentation, but not tested on a device yet. Check the map against the manufacturer's documentation before you write to a live device. Tested it? Report a test.

Communication

Protocol
Modbus RTU
Unit ID
47
Serial
9600 8N1 · 9600 baud, 8 data bits, none parity, 1 stop bit
Byte / word order
ABCD · bytes big-endian, words high first
Addresses
0-based

Connection values are defaults: the device may be configured differently.

Map

Registers
2 holding registers
Groups
2
Revision
1 · 2026-10-18
Added
()
File
maps/seeed-studio/s-zw-01a.mbmap.json · 2 KB · history
SHA-256
1cead27ba19bb8e37b6091e6ef5003f7e0bd3b9170b78c4b55e56c37620761c6
Manufacturer
seeedstudio.com

Sources and notes

Source: Seeed Technology, "UV Radiation Sensor (S-ZW-01) User Manual", V1.0

  • The reading is taken with FC03 from register 0; the manual's example answers 0x0050 = 80 W/m².
  • Factory settings: 9600 baud, 8 data bits, no parity, 1 stop bit. The manual gives 47 (decimal) as the default address of this sensor; its examples use address 1.
  • To change the address, write the new address to register 0x0100 with FC06 (the broadcast address 0xFE reaches a single sensor whose address is unknown). A changed address or communication setting takes effect after the sensor is powered off and on.
  • Range 0 to 200 W/m², resolution 1 W/m², supply 5 to 24 V DC.
AreaAddressNameTypeAccessUnitScaleDescription
HR 0 uint16 R W/m²
HR 256 uint16 RW Takes effect after the sensor is powered off and on

Help improve this map